Mident new brand dental air scaler cavitron MS-A3
Specifications
1. CE approved
2. Best Quality
3. Tips are made by materials imported from Sweden
4. 3 tips including: 1XS1, 1XS2, 1XS3.
Features
1. Easily attachable to any 2-hole and 4-hole borden handpiece connector. 2. Scope of application: applicable to scaling, and removal of calculus and stain. 3. Be sure to strictly control the air pressure. Over-high air pressure may damage or break tip in the mouth, and ca accident. 4. Even it's under standard air pressure range, the tip's amplitude sometimes may exceed 200 um, so please be sure give full attention while using and avoid harming patient. 5. Be sure not to use rusty tip, or it may bring the risk of tip break while using. 6. Autoclavable and sterilizable at 135℃ 7. Only when the interface air pressure is set to specified 280KPa, you can use this device. 8. The air pressure setting should not exceed 280KPa (2.9kgf/cm²) Introduction of dental scaler 3 Tips attached Technical Data
Standard Air Pressure 280KPa(2.9kgf/cm²) Air Consumption 25~35L/Min Frequency 5.600~5.850Hz 3 tips 1XS1, 1XS2, 1XS3 Weight 0.2KG Packing Size 18.5x8x4cm Dental air scaler main uses are stated as belows S1 - Ordinary size for scaling For the treatment in tooth distortion, root exposure in foretooth, and markedly tilted molar towards lingual side etc. S2 - Sickle-shape for scaling For the treatment in platy tartar in the mandible foretooth and surface along lingual lateral. S3 - Periodontal treatment For the treatment in subgingibal or supragingival calculus, and tarry attachments removal etc.